ClinMass® Applications



Immunosuppressants
Cyclosporine A, Tacrolimus, Sirolimus and Everolimus in Whole Blood – On-Line Analysis

The ClinMass® application for immunosuppressants represents a validated method for the therapeutic drug monitoring (TDM) of cyclosporine A, tacrolimus, sirolimus and everolimus. The sample preparation is based on a simple pre-treatment step, followed by a fully automated on-line sample analysis. The determination of the analytes is performed simultaneously with an injection interval of just 5 min. For calibration and measurement, we offer a 7 level whole blood calibrator set and two internal standards. For quality control whole blood controls in three different levels are available.
For cyclosporine A two additional levels (level IV, V) are offered.

  Fully automated on-line application
 Two internal standards
 7-Level whole blood calibrator set
 Whole blood controls in five levels

Test Data

Linearity: 
up to 4250 μg/l (Cyc), up to 200 μg/l (Tcr), up to 250 μg/l (Srl), up to 200 μg/l (Evr)
Lower detection limit: 5.7 μg/l (Cyc), 1.5 μg/l (Tcr), 0.74 μg/l (Srl), 1.2 μg/l (Evr)
Lower determination limit: 15.5 μg/l (Cyc), 4.0 μg/l (Tcr), 2.0 μg/l (Srl), 3.2 μg/l (Evr)
Intraassay precision: 3.7 % (Cyc), 2.8 % (Tcr), 3.3 % (Srl), 2.4 % (Evr)
Interassay precision: 5.1 % (Cyc), 4.3 % (Tcr), 3.9 % (Srl), 4.1 % (Evr)


LC-MS/MS Parameters

LC-Parameters

Pump 1: HPLC, flow rate 3 ml/min, water / methanol (90:10)
Pump 2: HPLC, flow rate 0.5 ml/min, methanol / ammonium acetate buffer (97:3)
Injection volume: 50 μl, Injection interval: 5 min
Switching valve: automatic switching valve, 6-port/3-channel
SPE column: Applied Biosystems Poros R1/20
Analytical column: Phenomenex Luna Phenyl-Hexyl


MS-Parameters

MS/MS-System: Applied Biosystems API 3000 Ion source: ESI positive

 Immunosuppressants LC-MSMS

  Back to top  


Metabolic Diseases
Homocysteine in Plasma


Homocysteine is as an important risk factor for vascular diseases. Elevated levels of homocysteine particularly have been shown to be associated with an increased risk of coronary heart disease and a higher likelihood of suffering a stroke.  The ClinMass® application for homocysteine represents a validated method for the quantification of total plasma homocysteine by tandem mass spectrometry. After an easy and fast two-step sample preparation (reduction, followed by precipitation of the protein matrix), the samples are rapidly analysed by the LC-MS/MS system within an injection interval of 2 min.  For calibration and measurement, we offer a plasma calibrator and an isotope labelled internal standard. For quality control plasma controls in two different levels are available.

  Simple and quick two-step sample preparation
  Injection interval: 2 min
 
Isotope labeled internal standard, plasma calibrator and plasma controls available

Test Data

Linearity: 0.6 - 92 μmol/l
Lower detection limit: 0.2 μmol/l, Lower determination limit: 0.6 μmol/l
Intraassay precision: 4.2 %, Interassay precision: 4.4 %


LC-MS/MS Parameters

LC-Parameters

Pump: HPLC, flow rate 0.7 ml/min, acetonitrile / water
Injection volume: 1 μl, Injection interval: 2 min
Analytical column: Supelcosil LC-CN


MS-Parameters

MS/MS-System: Applied Biosystems API 3000 Ion source: ESI positive


Sample Preparation

Homocystein in Plasma Sample Preparation for LC-MS/MS
Homocysteine in Plasma LC-MS/MS Application