CDT (Carbohydrate-Deficient Transferrin) in Serum - On-Line Analysis, Order No. 21100

 

This test is designed as a fast (injection interval: 9.5 min) and fully automated method for CDT analysis. A manual sample preparation is not required, but is performed on-line within the HPLC system. Equally to the CDT off-line kit, the test procedure closely directs to the recommendations of the IFCC-CDT-WG* for standardisation of CDT analysis.

For the transferrin detection a sensitive and cost-efficienct CDT detector is available (order no. FW3000).

*Working group of the IFCC for the international standardisation of CDT measurement.

 
  • Fully automatic method for CDT analysis


  • Test procedure closely directed to the "candidate reference method" of the IFCC-CDT-WG*


  • Injection interval: 9.5 min


Test Data

 

HPLC Parameters

Lower Detecrion limit: 0.1 %*
Lower determination limit: 0.3 %*
Intraassay precision: 4.5 %
Interassay precision: 3.9 %
*depending on the amount on total transferrin and on detector sensitivity
 
Pump 1,
Mobile Phases A and B:
binary gradient pump,
flow rate: 1.4 ml/min
Pump 2,
Mobile Phase C :
isocratic pump,
flow rate: 0.4 ml/min
Injection volume: 400 - 500 µl
Injection interval: 12 or 9.5 min*
  *The chromatograpic separation can be performed with a 12- or 9.5 min program
UV/VIS-Detector: 460 nm (e.g. RECIPE Fixed-
Wavelenght-VIS-Detector
for CDT)
HPLC-Thermostat: 40 °C

 

Sample Pretreatment

sample   preparation

 

Chromatograms (9.5 min program)

Native sample, normal range Native sample, pathological range
21000_chrom1   21000_chrom2
Disialotransferrin: 4.50 min, Trisialotransferrin: 5.17 min, Terasialotransferrin: 6.17 min,
Pentasialotransferrin: 7.13 min,
Hexasialotransferrin: 7.86 min
  Asialotransferrin: 3.02 min, Disialotransferrin: 4.52 min, Trisialotransferrin: 5.19 min, Terasialotransferrin: 6.21 min, Pentasialotransferrin: 7.15 min,
Hexasialotransferrin: 7.84 min


CDT (Carbohydrate-Deficient Transferrin) in Serum - On-Line Analysis, Order No. 21100

Ordering Information

Order No. Description Quantity
21100  ClinRep® Complete Kit for CDT in Serum for 300 assaysfor 300 assays 1 pce. 
 
  Kit Components (order no. 21100):  
21007 Mobile Phase A 2000 ml
within the kit: 3 x order no. 21007
21008 Mobile Phase B 2000 ml
21009 Mobile Phase C 2000 ml
within the kit: 2 x order no. 21009
21120 Vials for Sample Pretreatment with Diluting Solution 50 pcs.
within the kit: 6 x order no. 21120
  Manual 1 pce.
  Quick Reference 1 pce.
 
  Optional Components:  
21014 Test Solution 15 ml
 
  Start Accessories:  
21030 Analytical Column with test chromatogram 1 pce.
21032 Cartridge incl. 1 precolumn 1 pce.
FK5820 Stainless steel prefilter holder 1 pce.
FK7400 Inline-Filter (stainless steel sieve, free of dead volume) 1 pce.
 
  Accessories:  
21033 Precolumns 5 pce.
 
  ClinChek® Controls:  
21080 Serum Control, lyophil., Level I 10 x 1 ml
21081 Serum Control, lyophil., Level II 10 x 1 ml